OPERATION
Axio Examiner
Illumination and contrast procedures
Carl Zeiss
M60-2-0003 e 05/2012
63
(4) Centering the DIC slider
Only for DIC sliders for objectives with a parfocal length of 45 mm.
Figure 4-5 Axio Examiner.D1 with DIC equipment
−
The DIC slider (Figure 4-5/
1
) is inserted properly into the objective adapter.
−
The microscope adjusted for a bright field according to KÖHLER's rules.
−
The 10x objective (or the overview objective used) (Figure 4-5/
4
) is in working position.
−
The condenser turret has been swiveled into the bright field or into an unequipped position.
•
Swing the SÉNARMONT polarizer (Figure 4-5/
2
) into the optical path.
•
Turn the thumb wheel (Figure 4-5/
3
) at the SÉNARMONT polarizer to the 45° lock-in position.
•
Swing in the analyzer module at the reflector turret (Figure 4-5/
8
).
•
Remove one eyepiece from its eyepiece socket, and insert an auxiliary microscope.
•
A diagonal black strip of the DIC slider is seen when observing the visual field through the auxiliary
microscope.