9) The image position of
end-face gap leans to one
side of screen. (The
function can follow the
arc position
automatically, if too
much, please adjust it.
1.
Too much
sediment on the
electrode.
2.
Electrode deflects
because of
concussion.
3.
The lens leans to
unusual position.
4.
When testing
current, open the
windshield too
early or touch one
side fiber, this is
because of wrong
judgment of arc
position.
1.
Enter the menu and
clean electrode
several times.
2.
Close machine and
restart after several
minutes, then do arc
test at least three
times.
3.
If deflection isn’t
too much, do not
adjust it.
4.
Contact us due to
special condition.
10) After cleaning V-
groove, fiber can’t be put
in it well and the image
is unclear.
1.
The groove on the
fiber holder and
V-groove don’t lie
in same line or
same flat.
2.
The tail fiber can
not be pressed
tightly.
1.
Open the board,
push the groove
toward to near V-
groove, check
whether the two
grooves lie in same
line and same flat,
then manage it.
2.
Reload fiber.
3.
If it doesn’t work,
contact us.
All problems are
caused by inner
circuit fault
probably.
If you couldn’t
solve the
problems,
please sent it
to us with
detailed
explanation.
11) Normal setting gap
and calibration, but
electrode doesn’t arc
during the fusion.
1.Set the current
parameter to zero or
select the program
without setting
parameter.
2. The storage battery
doesn’t work well.
3.When press the
“RUN” and the
machine is in mode of
auto splice,” HEAT”
is pressed by mistake.
1.Enter the menu, check
whether the parameter
setting is normal, if not,
change it to normal
parameter. If this
doesn’t work, please
send it to us to change
the battery.
2. Change the battery
3. Don’t press” HEAT”
when the “START” is
pressed and the machine
is in mode of auto
splice.
12) There is spark in the
electrode room or
electrode arcs to near
metal.
1.The environment is
too damp.
2.Elecrode is loose or
deformed.
1.Change to dry
environment.
2.Open windshield to
check electrode is loose
or deformed. If it
doesn’t work, contact
us.