22
5>
Press “ENT” to enter maintenance menu;
6>
Press “▲” or “▼” to make cursor “
→
” point to “CLR Electrode”;
7>
Press “ENT” to make machine arc automatically, do those two or three
times. To stabilize arc current and clean electrode, the high temperature
will boil off the impurity on the electrode, until arc is stable and the
sound of “hiss” disappear.
8>
Press “MENU” to exit menu in turn.
Notice:
1)
Because top of electrode is very weak, to avoid damage of electrode and
unstable arc, do not touch it by hard things.
2)
During the process of maintenance, to avoid damage of instrument, do
not arc large current for a long time.
10.3
Cleaning V-groove
The driving range of adjusting core’s direction is only a little micron, fiber’s
picture will departure normal position and couldn’t level if any things exists.
You should clean the V-groove timely as following operations:
1>
Open the windshield;
2>
Open the fiber holder and fiber clamp;
3>
Using the special tool to push several times at the same direction.
☆
Notice:
To avoid damage of V-groove and instrument, do not clean V-groove
with hard thing(knife) and energize on it.
10.4
Cleaning lens of microscope
Fusion splicer adopts the working mode of disposal pictures. It is
necessary to keep the lens of microscope cleaning. The operations are as follows:
1>
Open the windshield, under the electrode you will find two lens.
2>
If any thing exists on lens, dispose it with proper instrument. To avoid of
damage, do not touch lens with hard thing.
3>
Wipe lens with a cotton swab soaked with absolute alcohol or air pouch.
4>
Wipe lens tenderly with dry cotton swab until it is clean.
10.5
Maintenance
This company guarantees to keep fusion splicer in good repair. Do not
open box to repair if any failure exists. To resolve questions in time, please contact
with our technology department at once.