9.2
Setting working parameter
Working parameters of
Fusion Splicer
include heating time, surface-angle, tension test, etc.
The operations are as follows:
1>
Press “MENU” to enter the first class menu;
2>
Press “▲” or “▼” to make cursor “
→
” point to working parameter to be selected.
3>
Press “◄” or “►” to select needed working parameter.
4>
Press “ENT” to confirm.
5>
Press “MENU” to exit the first order menu.
●
Notice:
1)
The value of “Setting end-face” should be 6 usually under the premise
of fusion automatically.
2)
You can adjust “Heating time” freely, the heating control of KL-280 very
flexible, it can be controlled by keyboard or reckon through time.
3)
“Fusion fiber” must be accord with actual spliced fiber.
10.
Maintenance
10.1
Replacing the electrode
Firstly make sure fusion splicer is turned off, the take off protective cover of electrode room,
loosen the screw and take out up-electrode. Then take out down-electrode in the same way. The
installation’s order of new electrode is opposite to strip. Gap between two tops of electrode should
be within 2.6
±
0.2mm and they should be symmetrical to fibers. Usually, do not adjust electrode,
and to avoid damage of electrode do not touch the electrode’s top in the whole process, prevent
electrode dropping into the machine. After replacing electrode, you should aging electrode, test the
current until it is steady and do the arc test , after “Current is OK” is showed 3 times, user can
continue to splice, steps are as following:
Press “MENU” to enter first class menu,
1)
Press “▲” or “▼” to make cursor to “Maintenance ”
2)
Press “ENT” to enter it
3)
Press “▲” or “▼” to make cursor to “electrode aging”
4)
Press “ENT” to achieve arcing, arcing count is shown at the bottom of displayed, 20
times of arcing is shown in one time of “Electrode aging”, after buzzer bleats twice,
electrode aging is finished. User can know the electrode situation according to steady of
image and sound.
Notice:
1)
Shut off the power of machine before replacing electrode.
2)
Make sure that arc parameter had been setup efficiently in the arc process
before testing the arc current.
10.2
Maintenance of electrode
After extended use, the top of electrode will produce dust; dust will make
arc unsmooth. When “hiss” sounds, you should clean electrode. Best
to clean electrode timely as following operations:
1>
Open the machine ;
2>
Clean electrode’s top tenderly with alcohol tampons;
3>
Press “MENU” to enter the first class menu;
4>
Press “▲” or “▼” to make cursor “
→
” point to “Maintenance”;