www.youfibercable.com
Email:[email protected] MSN:[email protected]
24
displayed on screen and after a while “Reload fiber” is displayed on screen.
Solutions:
1) Raise the windshield, left and right clamps. The checking steps as follows:
2) Check whether broken fiber exists.
3) Check whether cleaved length is too short.
4) Check whether fiber matches with V-groove.
6> When calibrating automatically, fiber moves up and down, the screen’s
display stops at “Adjust core”.
Solutions:
1) Check whether the position warp between X and Y directions’ surface is less
than1cm. (screen displays size) If not, return KL-280 factory to fix, otherwise
operating as follows:
2) Check whether bare fiber is clean.
3) Clean the dust of V-groove.
7>The fiber can be reset as normal, but when setting gap, “Setting gap” is
displayed on screen and after a while “Bad left fiber’s surface” is displayed.
Solutions:
1)
Observe fiber picture in screen, if the quality of left fiber’s surface is bad
,
you can
remake surface and try again.
2)
Observe fiber picture in screen, if the quality of left fiber’s surface is good,
maybe the value of “Setting surface” is too small, if you want to keep on working
you can increase it.
3)
After “Bad left fiber’s surface” is displayed, the screen becomes dark;
Make sure that windshield is covered well.
Raise windshield, check whether the pipe is dent, and then manage it.
8> The fiber can be reset as normal, but the time of auto-fusion is too long.
Solutions: Enter the menu of arc parameter; check whether the setup is
efficient.
9>During arc test, the position of fiber’s gap leans to one side of screen.
Solutions: When the machine tests the arc, it will adjust the position of current and
arc. If the arc on the apparent of electrode were asymmetric, the arc position
would change. If the change was not too much, you can ignore it, otherwise please
mange it:
1)
Enter the menu of maintenance, and then operate “Clean the electrode” several times.
10> When arc splicing, do not use the program (1~5).
Solutions: Because of electrode’s aging, biggish change takes place in opposite
position between fiber and arc or operating environment.
1)
The condition of electrode’s aging.
Operate “Clean the electrode” if the top of electrode is well.
Replace electrode if the top of electrode is bad. (Consult 10<Maintenance>)
2)
Big change takes place in operation environment. Operations are as follows:
1>
Do arc test until “Moderate current” is displayed three to five times.