14
Table 4
Shape
Reason
Measure
bubble
1.
There is dust on the end-face
2.
Dew
3.
Bad fiber’s end-surface
4.
Too small arc current
1.
Remake end-surface.
2.
Adjust program or current.
(
Optimization of arc
parameter)
broken
thin
diameter
1.
Too large arc current
2.
Too small push speed
3.
Little fusion push
4.
Feed blocked
1.
Adjust program or current
parameter.
2.
Clean V-groove.
(
Optimization of arc
parameter)
black line
Notice: Black line will appear
on fiber’s joint with fluorine,
it has no effect to transport
characteristic.
Press “RUN” to arc again
and observe.
6.4
Tension test
KL series fusion splicers supply the function of tension test.
If you need this function, when you enter into “MENU”, please open this function, and after each
splicing, machine will do tension test.
After you start the machine, the screen displays as left picture:
6.5
Protection of fusion point
6.5.1
Taking out spliced fiber
Raise windshield and fiber-collect in turn, take out spliced fiber, do not touch electrode.
6.5.2
Moving splice protection sleeve
Move splice protection sleeve that covered around fiber to joint, make sure that length
of two fiber’s coat is equal in splice protection sleeve. Consult figure 8(a).
6.5.3
Heating splice protection sleeve
You can base need to select “if test
tension”, then operate. After splicing
the screen displays as right picture.
Press “ENT” to make KL-260C to test
tension automatically.
If you don’t want to test tension, then press
“RESET” to reset system after taking out
spliced fiber or do not select “Test tension”.
09-03-10 15:30
AUTO (SM)
Proof Ok
Estibate Loss: 0.01dB