<9. Parameter Summary>
9-1
IM 01C50E01-01EN
9. Parameter Summary
Table 9.1
Parameter Summary
Object
ID
Attribute
ID
Label
Description
Default value Handling
1.
UAPMO
block
1
Version Revision
Indicates the application revision of YTA. This revision changes
when the application software is downloaded.
---
R
10
Static Revision
Indicates the revision level of the fixed parameters of UAP.
Used, for example, to check whether parameters have been
changed.
0
R
64
Identification
Number
Indicates the vendor ID, model ID, and revision of the device.
---
R
65
CTS Version
Indicates the version of the communication stack test system
(CTS).
0
R
66
ITS Version
Indicates the version of the interoperability test system (ITS).
0
R
67
Diagnostic Status
Indicates the diagnostic results of the device based on the
NAMUR NE107
*1
model.
Setting Enable diagnostic status configuration in UAP Option to
Enable allows turning OFF and ON the display of the diagnostic
results for each summary, and changing Categorize. For
Categorize at the time of shipment, refer to tables 8.3 and 10.2.
---
R
68
UAP Option
Allows setting the Diagnostic Status and write protection of UAP.
Software write protect
1: On, 0: Off
Enable hardware write protect
1: Enable, 0: Disable (default)
Enable diagnostic status configuration
1: Enable, 0: Disable (default)
The following table shows the relationship between the
hardware write protection and software write protection.
Enable
hardware
write
protect
Hardware
write
protect
Software write
protect
Write
protect
Disable
Off or On
Off
No
Disable
Off or On
On
Protected
Enable
Off
Off or On
No
Enable
On
Off or On
Protected
0
W
69
Diagnostic Switch Allows setting On/Off for each summary of Diagnostic Status
when Enable diagnostic status configuration in UAP Option is
set to Enable.
On
W
70
Diagnostic
Configuration
Allows Categorize for each summary of Diagnostic Status when
Enable diagnostic status configuration in UAP Option is set to
Enable.
Refer to Table
9.2.
W
102
Diagnostic Status
Detail
Detailed information on Diagnostic Status. For Categorize at the
time of shipment, refer to Table 9.2.
Refer to Table
9.2.
R
103
Energy Left
Indicates the number of days of remaining battery life assuming
ambient temperature condition as 23 degrees Celsius. The unit
is day.
---
R
104
Reset Energy Left Resets the remaining battery power calculation to restore it to
a remaining battery power calculation which is based on new
batteries.
0 = Continue
1 = Reset
0 (reading
value is
always 0)
W
105
Power Supply
Status
Indicates remaining battery life and power supply of device
0 – line powered
1 – battery powered, greater than 75% remaining capacity
2 – battery powered, between 25% and 75% remaining
capacity
3 – battery powered, less than 25% remaining capacity
---
R
Note: TSAP ID is 2.