3-12
IM WT5000-02EN
Procedure Using the Setup Menu
1.
Tap the
Setup
icon
, or press
MENU
under SETUP.
2.
Tap
Computation/Output
tab. A computation and output settings overview screen appears.
Pressing
ESC
closes the overview screen.
Setting Measurement Equations (Measure)
3.
Tap
Measure
.
An equation setup screen appears.
Computation/Output tab
Formula screen
Set equations.
Set the apparent power equation
(Urms*Irms, Umean*Imean, Udc*Idc,
Umean*Irms, Urmean*Irmean).
Set the apparent power and reactive power
computation type (Type 1, Type 2, Type 3).
Set the corrected power
standard to be applied (Type1,
Type2).
Set the coefficients (0.0001 to
9.9999).
When Select standard is Type1,
set coefficients P1 and P2.
Note
You can also display the computation/output settings overview screen by moving the cursor on the
Computation/Output tab using the arrow keys and then pressing SET.
3.3 Displaying Apparent Power, Reactive Power, and Corrected Power