5-4
IM WT5000-02EN
Procedure Using the Setup Menu
1.
Tap the
Setup
icon
, or press
MENU
under SETUP.
2.
Tap
Computation/Output
tab. A computation and output settings overview screen
appears.
Pressing
ESC
closes the overview screen.
Switching the Displayed Items (Items)
You can switch the measured value (measurement function) shown in the screen.
3.
Tap
Display
.
A display format setup screen (Numeric/Graph) appears. For details, see section 3.1.
4.
Tap
Items
. The following screen appears.
Closes the menu
Set the displayed item.
Set the displayed harmonic items. Scroll the display list, and select the displayed items from
the following available items.
• Harmonic order k (Harmonics(k) Group):
U (k)
(rms voltage of harmonic order k),
I (k)
(rms current of harmonic order k),
P (k)
(active power of harmonic order k),
S (k)
(apparent power of harmonic orderk),
Q (k)
(reactive power of harmonic order k),
λ(k)
(power factor of harmonic order k),
Φ(k)
(phase difference between the voltage and current of harmonic order k),
ΦU(k)
(phase difference between harmonic voltage U(k) and the fundamental wave U(1)),
ΦI(k)
(phase difference between harmonic current I(k) and the fundamental wave I(1)),
Z (k)
(load circuit impedance),
Rs (k)
(load circuit series resistance),
Xs (k)
(load circuit series reactance),
Rp (k)
(load circuit parallel resistance),
Xp (k)
(load circuit parallel reactance),
Uhdf (k)
(harmonic voltage distortion factor),
Ihdf (k)
(harmonic current distortion factor),
Phdf (k)
(harmonic active power distortion factor),
• Harmonics (Harmonics Group):
Uthd
(total harmonic voltage distortion),
Ithd
(total harmonic current distortion),
Pthd
(total harmonic active power distortion),
Uthf
(voltage telephone harmonic factor),
Ithf
(current telephone harmonic factor)
Utif
(voltage telephone influence factor),
Itif
(current telephone influence factor),
hvf
(harmonic voltage factor),
hcf
(harmonic current factor),
K-factor
(K factor),
ΦUi-Uj
(phase difference between the fundamental voltage of element i, Ui(1),
and the
fundamental voltage of element j, Uj(1)),
ΦUi-Uk
(phase difference between Ui(1) and the fundamental voltage of element k, Uk(1)),
ΦUi-Ii
(phase difference between Ui(1) and the fundamental current of element i, Ii(1)),
ΦUj-Ij
(phase difference between Uj(1) and the fundamental current of element j, Ij(1)),
ΦUk-Ik
(phase difference between Uk(1) and the fundamental current of element k, Ik(1)),
• PLL source frequency (Frequency group):
FreqPLL1
(PLL1 frequency),
FreqPLL2
(PLL2 frequency)
Set the harmonic order (Total, 0, 1 to 500).
Procedure Using the Menu Icons
You can also use the menu icons shown on the right side of the screen to switch the displayed
items.
1.
Tap the
Display
menu icon
. A Display menu appears in the sub menu area on the
right side of the screen.
By tapping the displayed items, you can specify the same settings as when using the screen
explained earlier.
Note
For details on the Display menu, see page iv.
5.2 Displaying Harmonic Measurements (numeric display)