16.2 Configuring the D/A Output (option)
► “D/A Output (D/A Output, option)” in the features guide
This section explains operating procedures using the following setup methods.
• Procedure Using the Setup Menu (see chapter 1)
Procedure Using the Setup Menu
1.
Tap the
Setup
icon
, or press
MENU
under SETUP.
2.
Tap
Computation/Output
tab. A computation and output settings overview screen appears.
Pressing
ESC
closes the overview screen.
Configuring the D/A Output (D/A Output)
3.
Tap
D/A Output
. A DA output setup screen appears.
Computation/Output tab
Configure the
D/A output
items.
D/A output signal name
For details on the connector pinout and the D/A
output signal assignment, see section 4.5 in the
Getting Started Guide, IM WT5000-03EN.
Output items
This display changes according to
the Function, Element/Σ, and Order
settings.
Set the measurement function (None, other
functions—
for details on the various measurement
functions, see “Items That This Instrument Can
Measure” in the Features Guide).
Switches the channel display group
Ch 1-Ch 5: Displays Ch1 to Ch5,
Ch 6-Ch 10: Displays Ch 6 to Ch 10
Ch 11-Ch 15: Displays Ch11 to Ch15
Ch 16-Ch 20: Displays Ch 16 to Ch 20
Set the rated time of integrated
D/A output
(00000 h : 00 min : 00 s to
10000 h : 00 min : 00 s).
Set the harmonic order (Total, 0 to 500).
You can set this setting when the
measurement function includes a harmonic
order.
Select the D/A output range mode (Fix, Manual).
Set the maximum and
minimum values of the
range (–9.999T to 9.999T).
These can be set when
range mode is set to
Manual.
Selects the input element or wiring unit to be configured
• Input element options
When the displayed item is set to something other than motor
evaluation: Element1 to Element7
When the displayed item is set to motor evaluation: Motor1 to Motor4
• Wiring unit options: ΣA, ΣB, ΣC
Note
You can also display the computation/output settings overview screen by moving the cursor on the
Computation/Output tab using the arrow keys and then pressing SET.