2-34
IM WT5000-02EN
2.13 Setting Averaging
► “Averaging (Averaging)” in the features guide
This section explains operating procedures using the following setup methods.
• Procedure Using the Setup Menu (see chapter 1)
Depending on the measurement mode or the method of computing measurement values (computing
method), the settings may be different from the description in this section, the settings may be invalid,
or the settings may not be configurable. For details, see the features guide.
Procedure Using the Setup Menu
1.
Tap the
Setup
icon
, or press
MENU
under SETUP.
2.
Tap
Computation/Output
tab. A computation and output settings overview screen appears.
Pressing
ESC
closes the overview screen.
Setting Averaging (Averaging)
3.
Tap
Update Rate/Averaging
.
A data update interval/averaging setup screen appears.
Computation/Output tab
Update Rate/Averaging button
Averaging screen
Turns averaging on and off
Set the averaging type (Exp, Lin).
Set the attenuation constant or
average count.
• When Type is set to Exp:
Attenuation constant (2 to 64)
• When Type is set to Lin: Average
count (8 to 64)
Update Rate screen
(see section 2.10)
Note
You can also display the computation/output settings overview screen by moving the cursor on the
Computation/Output tab using the arrow keys and then pressing SET.