6.2 Codes and Corrective Actions
Error Codes
An error code may appear on the screen during operation. This section describes the
meanings of the messages and their corrective actions. If the corrective action indicates
servicing, contact your nearest YOKOGAWA dealer for repairs.
In addition to the error codes below, there are Acquisition Software error messages.
These are described in the
Acquisition Software User’s Manual IM720120-61E
.
Status Messages
Type Code Description
Corrective Action
Reference
Section
M
52
Auto setup is in progress.
–
IM720120-61E
M
53
Auto setup is complete.
–
M
58
Undoing auto setup.
–
M
59
Auto setup has been undone.
–
M
66
Strain balancing in progress.
–
M
67
Strain balancing is complete.
–
M
69
Calibrating (for manual calibration).
–
M
70
Calibration is complete (for manual calibration). –
M
70
Internal hard disk formatting is complete.
–
M
78
Undoing initialization.
–
M
79
Initialization has been undone.
–
M
81
The total sync cable length exceeds the
maximum allowable length (10 m), and the sync
clock delay tolerance has been exceeded.*
Connect the SL1000s so that the total length
of all sync cables is less than or equal to 10 m.
Section 5.7
M
88
Initializing settings.
–
IM720120-61E
M
89
Initialization is complete.
–
* The message appears during synchronous operation on all slave devices whose total cable
length from the master unit exceeds 10 m.
Execution Errors
Type Code Description
Corrective Action
Reference
Section
E
703
Cannot undo because auto setup nor
initialization is not in progress.
–
IM720120-61E
E
713
Auto calibration failed.
Check the input signal at the specified channel.
E
714
Measurement start was executed even though
the unit is not in sync operation mode.
Start the Acquisition Software, set the unit
to sync operation mode, and then start
measurement.
E
724
Strain balancing failed.
Check the connection at the specified channel.
E
755
The time base cannot be set to external clock
for Envelope and Box Average.
Switch to internal clock.
E
777
Range-over occurred during shunt calibration.
Increase the measuring range and execute the
calibration again.
Configuration Errors
Type Code Description
Corrective Action
Reference
Section
E
806
Cannot change settings while GO/NO-GO is in
progress.
Abort GO/NO-GO and then change the
settings.
IM720120-61E
E
821
Cannot change the trigger delay when using
external clock.
Switch to internal clock.
E
877
Cannot change settings because there are too
many display channels for the current record
length.
Shorten the record length.