3-4
IM 720120-01E
• Place the instrument in a horizontal position as shown below.
• If you are installing the instrument on a slippery surface, attach the rubber feet (two
pieces) to the rear feet on the bottom panel.
Installation Conditions
Install the instrument in a place that meets the following conditions:
Well-Ventilated Location
There are vent holds on the right and bottom panels of this instrument. In addition, there
are exhaust holes for the cooling fan on the rear panel. To prevent internal overheating,
allow for enough space around the instrument (see the figure below) and do not block
the vent and inlet holes.
5 cm
or
more
5 cm or more
10 cm or
more
5 cm
or
more
HIGH-SPEED DA
TA ACQUISITION UNIT
SLOT
1
SLOT
2
3
SLOT
SLOT
4
5
SLOT
SLOT
6
7
SLOT
SLOT
8
EXT C
LK IN
POWER
TRIG
'D
START
/
STOP
DISPL
AY
PROBE
POWER
( 12
V DC
)
When you are installing modules or connecting cables, allow extra working space.
3.2 Installing the Instrument