5-10
IM 720120-01E
To remove a signal wire, press down on the top section of the screwless terminal with a
flat-blade screwdriver and pull the wire out.
Note
Be sure to hold the SL1000 firmly while pressing down on
the top section of the screwless
terminal
with a flat-blade screwdriver.
Use the following signal wires for the GO/NO-GO output.
• Signal wire thickness (single wire):
ϕ0.4 to 1.0 (AWG26 to 18)
(stranded wire):
0.3 to 0.75 mm
2
(AWG22 to 20)
• Recommended length of stripped wire: 10 mm
NO-GO OUT Signal
When the judgment result is NO-GO (fail), the output signal level (TTL level) temporarily
changes from high to low.
GO OUT Signal
When the judgment result is GO (pass), the output signal level (TTL level) temporarily
changes from high to low.
Circuit Diagram of the GO/NO-GO Output
GO/NO-GO
120
Ω
+5 V
4.7 k
Ω
GO/NO-GO OUT Output Timing
Waveform
acquisition
Waveform
acquisition
Result
Judge
Judge
Result
NO-GO OUT
GO OUT
Output timing
EXEC
5 ms
or more
Remains at low output until the next
determination can be accepted.
If you specified an action that is carried
out when the condition is met, the time
is extended until the action is finished.
5.6 Connecting the GO/NO-GO Output Terminals