Making Preparations for Measurements
3-5
IM 720120-01E
1
2
3
4
5
6
7
App
Index
Ambient Temperature and Humidity
Use the instrument in the following environment.
• Ambient temperature: 5 to 40°C
• Ambient humidity:
20 to 85%RH, no condensation.
Note
• To ensure high measurement accuracy, operate the instrument in the 23 ± 5°C temperature
range and 55 ± 10% RH.
• Condensation may occur if the instrument is moved to another place where the ambient
temperature is higher, or if the temperature changes rapidly. If condensation occurs, allow
the instrument adjust to the ambient temperature for at least an hour before using the
instrument.
Do Not Install the Instrument in the Following Places
• Outdoors.
• In direct sunlight or near heat sources.
• Where the instrument is exposed to water or other liquids.
• Where an excessive amount of soot, steam, dust, or corrosive gas is present.
• Near strong magnetic field sources.
• Near high voltage equipment or power lines.
• Where the level of mechanical vibration is high.
• In an unstable location.
Storage Location
• We strongly recommend you store the SL1000 in an environment with a temperature
between 5 and 40°C and a relative humidity between 20 to 80%RH.
• When storing the SL1000, avoid the following locations.
• In direct sunlight.
• Where the temperature is 60°C or higher.
• Where the relative humidity is 80% or more.
• Near heat sources.
• Where the level of mechanical vibration is high.
• Where corrosive or explosive gas is present.
• Where an excessive amount of soot, dust, salt, and iron are present.
• Where water, oil, or chemicals may splash.
3.2 Installing the Instrument