2.1 System Configuration and Block Diagram
System Configuration
Signal input
Measured item
• Module
1
• Probe
2
PC
• External clock input
• External trigger input
• Sync input
• GO/NO-GO output
• Trigger output
• Alarm output
• Sync output
• USB interface
• Ethernet interface (option)
1 • High-Speed 100 MS/s, 12-Bit Isolation Module
• High-Speed 10 MS/s, 12-Bit Isolation Module
• High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module
• High-Speed 10 MS/s, 12-Bit Non-Isolation Module
• High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS)
• High-Voltage 1 MS/s, 16-Bit Isolation Module (with RMS and AAF)
• Temperature, High Precision Voltage Isolation Module
• Universal (Voltage/Temp.) Module
• Universal (Voltage/Temp.) Module (with AAF)
• Strain Module (NDIS)
• Strain Module (DSUB, Shunt-Cal)
• Acceleration/Voltage Module (with AAF)
• Frequency Module
2 • Isolated probe
• Current probe
• Passive probe
Internal hard disk (option)
• FTP server
• Setup data
• Measured data
Trigger sources and alarm sources can be combined between units (using AND or OR logic).
• External clock
input
• External trigger
input
• Trigger output
• Alarm output
Master
UNIT-ID=0
Slave #1
UNIT-ID=1
Slave #2
UNIT-ID=2
Slave #7
UNIT-ID=7
Signal input
(Trigger source,
Alarm source)
Synchronous Operation
Signal input
(Trigger source,
Alarm source)
Signal input
(Trigger source,
Alarm source)
Signal input
(Trigger source,
Alarm source)
SYNC
OUT
SYNC
IN
SYNC
OUT
SYNC
IN
SYNC
OUT
SYNC
IN
During synchronous operation, the master unit’s GO/NO-GO output is invalid. The slave
units’ external clock input, external trigger input, trigger output, alarm output, and GO/
NO-GO output are invalid.
The external clock input and external trigger input synchronize with the signals from the
master unit.
Trigger sources and alarm sources are combined using AND or OR logic between units,
and are transmitted from the master unit’s trigger output and alarm output.
Chapter 2
Explanation of Functions