7-6
IM 720120-01E
Item
Specifications
701255: 10 MS/s, envelope mode, 500 mV range, no input filter, with the tip of the probe
(701940 (10:1)) shorted
701267: 1M S/s, envelope mode, 2 V range, no input filter, with the tip of the measuring lead
shorted
720268: 100 kS/s, envelope mode, 2 V range, no input filter, with the tip of 701904 shorted
701261: 100 kS/s, envelope mode, 50 mV range, no input filter, with the end of the 3-m
twisted-pair cable shorted
701262: 100 kS/s, envelope mode, 50 mV range, no input filter, with the end of the 3-m
twisted-pair cable shorted
701265: 100 kS/s, envelope mode, 1 mV range, no input filter, with the end of the 3-m twisted-
pair cable shorted
720266: 100 kS/s, envelope mode, 1 mV range, no input filter, with the end of the 3-m twisted-
pair cable shorted
701270: 100 kS/s, envelope mode, 500 μSTR, gauge factor: 2.0, no input filter
701955 bridge voltage: 2 V
701956 bridge voltage: 10 V
701271: 100 kS/s, envelope mode, 500 μSTR, gauge factor: 2.0, no input filter
701957 bridge voltage: 2 V
701958 bridge voltage: 10 V
701275: 100 kS/s, envelope mode, 500 mV range, no input filter, with the tip of the probe
(700929 (10:1)) shorted
701281: 1 MS/s, envelope mode, Frequency, 1 Hz range, no input filter
User (±1 V, threshold level: 0 V, hysteresis: ±1%), with the tip of the probe (700929
(10:1)) shorted
720281: 1 MS/s, envelope mode, Frequency, 1 Hz range, no input filter
User (±1 V, threshold level: 0 V, hysteresis: ±1%), with the tip of the probe (700929
(10:1)) shorted
720210: 100 MS/s, envelope mode, 1 V range, no input filter, with the tip of the probe (700929
(10:1)) shorted
720211: 100 MS/s, envelope mode, 1 V range, no input filter, with the tip of the probe (700929
(10:1)) shorted
Test Item
1. Electrostatic discharge
EN61000-4-2
Air discharge: ±8 kV, contact discharge: ±4 kV, criteria B
2. Radiated immunity
EN61000-4-3
80 MHz to 1 GHz, 1.4 GHz to 2 GHz, 10 V/m Criteria A
2 GHz to 2.7 GHz, 3 V/m Criteria A Criteria A
3. Conducted immunity
EN61000-4-6
3 V, criteria A
4. Electrical fast transient/burst
EN61000-4-4
Power line: ±2 kV, signal line: ±1 kV, criteria B
5. Power frequency magnetic field EN61000-4-8
30 A/m, 50 Hz, criteria A
6. Surge immunity
EN61000-4-5
Between lines: ±1 kV, common: ±2 kV, criteria B
7. Voltage dip and interruption
EN61000-4-11
1 cycle, both polarities, 100%, criteria A
10 cycle, both polarities, 60%, criteria A
25 cycle, both polarities, 30%, criteria A
250 cycle, both polarities, 100%, criteria C
Definitions of criteria A, B and C
Criteria A: Aforementioned “Influence in the immunity environment” is met during the test.
Criteria B: This apparatus continues to operate without hang-up or falling into uncontrollable
conditions during the test. No change of actual operating state or stored data is
allowed.
Criteria C: When the noise is applied, temporary degradation or less of function is observed
and to recover the normal condition, an operator intervention or system reset is
required.
Environmental standard
Compliant Standard
5
EN50581 Monitoring and control instruments
Approved (720120, 720211, 701250, 720250, 701251, 701255, 701267, 720268, 701261,
701262, 701265, 720266, 701270, 701271, 701275, 701281, 720281, 720210, 700986,
700987, 701955, 701956, 701957, 701958)
1. If the 701260 module is inserted into the SL1000, it will not comply with safety standard EN61010-1:2010 or EN61010-2-030:2010.
2. The Overvoltage Category (Installation Category) is a value used to define the transient overvoltage condition and includes the
impulse withstand voltage regulation. I applies to electrical equipment that is powered by a circuit with overvoltage control. II
applies to electrical equipment that is powered through a fixed installation such as a wall outlet wiring to a switchboard.
7.7 General Specifications