3-8
IM 720120-01E
Types of Input Modules
The following types of input modules are available.
• High-Speed 100 MS/s, 12-Bit Isolation Module:
720210 (HS100M12)
• High-Speed 100 MS/s, 12-Bit Isolation Module:
720211 (HS100M12)
• High-Speed 10 MS/s, 12-Bit Isolation Module:
701250 (HS10M12)
• High-Speed 10 MS/s, 12-Bit Isolation Module:
720250 (HS10M12)
• High-Speed High-Resolution 1 MS/s, 16-Bit Isolation Module: 701251 (HS1M16)
• High-Speed 10 MS/s, 12-Bit Non-Isolation Module:
701255 (NONISO_10M12)
• High-Voltage 100 kS/s, 16-Bit Isolation Module (with RMS): 701267 (HV (with RMS))
• High-Voltage 1 MS/s, 16-Bit Isolation Module (with AAF, RMS): 720268 (HV (AAF, RMS))
• Universal (Voltage/Temp.) Module:
701261 (UNIVERSAL)
• Universal (Voltage/Temp.) Module (with AAF):
701262 (UNIVERSAL (AAF))
• Temperature, High Precision Voltage Isolation Module: 701265 (TEMP/HPV)
• Temperature, High Precision Voltage Isolation Module
(low noise type):
720266 (TEMP/HPV)
• Strain Module (NDIS):
701270 (STRAIN_NDIS)
• Strain Module (DSUB, Shunt-Cal):
701271 (STRAIN_DSUB)
• Acceleration/Voltage Module (with AAF):
701275 (ACCL/VOLT)
• Frequency Module:
701281 (FREQ)
• Frequency Module:
720281 (FREQ)
Precautions to Be Taken When Installing or Removing Input Modules
Switching the installed input module with a different module and turning ON the power
initializes the settings on that channel. If you want to save the settings, use the SL1000
Acquisition Software to specify the save destination medium and save the settings. For
details, see section 8.2 in the
Acquisition Software User’s Manual IM720120-61E
.
CAUTION
Insert or remove an input module along the guide straight and slowly. Inserting or
removing a module at an angle can cause failure and damage.
French
ATTENTION
Insérer ou retirer tout droit et lentement un module d’entrée le long du guidage.
Le fait d’insérer ou de retirer un module de manière inclinée risque de causer une
défaillance et des dégâts.
Installation Procedure of Modules
1.
Check that the power switch on the rear panel of the instrument is OFF.
2.
Check the slot number indicated above the slots for installing the input modules
on the front panel of the instrument. Then, install the module along the guide.
Holding the handles on the top and bottom of the input module, press firmly
until it clicks in place. If there is a cover panel on the slot in which to install the
module, remove the cover panel, first.
3.
Make sure to screw the two locations, top and bottom, of the input module firmly
into place (screw tightening torque: 60 N•cm).
HIGH-SPEED DA
TA ACQUISITIO
N UNIT
SLOT
1
SLOT
2
3
SLOT
SLOT
4
5
SLOT
SLOT
6
7
SLOT
SLOT
8
EXT CLK
IN
POWER
TRIG’D
START/
STOP
DISPL
AY
PROB
E POW
ER(
12 V
DC
)
Guide
Attachment
screws
3.3 Installing Modules