![YOKOGAWA SC202G User Manual Download Page 95](http://html1.mh-extra.com/html/yokogawa/sc202g/sc202g_user-manual_3799615095.webp)
IM 12D08B02-01E
Appendix 3-7
3/3
QIS 12D08B02-21E
Table 3
Message display
Decade box 2 resistance
Data display
RES.1
10
:
10.00
±0.05
:
RES.2
100
:
100.0
±0.5
:
RES.3
1 k
:
1.000
±0.005
k
:
RES.4
10 k
:
10.00
±0.05
k
:
RES.5
100 k
:
100.0
±0.5
k
:
Figure 1 Testing Circuit and Test Equipment
SC202S
SUPPLY SENSOR
+
-
G
11
12
14
13
15
16
24V DC
+
100
:
DC source
+
-
-
DC
Decade
Box 1
Note 1: Cable connected to sensor input should be conductivity detector cable of
length 2.1 ±0.1 m. Connect pins 13 and 14, also 15 and 16, to Decade box 2
terminals. Use shielded cable, and connect shield to pin 14.
Note 1
Decade
Box 2
Ground
Milli-
ammeter