IM 12D08B02-01E
Appendix 3-5
13-2. SC202S 2-Wire Conductivity Transmitter
Quality
Inspection
Standards
SC202S
2-Wire Conductivity Transmitter
QIS 12D08B02-21E
1st Edition: Mar. 2007(YK)
1. Scope
This inspection standard applies to the SC202
□
2-Wire Conductivity Transmitter.
2. Inspection
Items
2.1 Insulation resistance test
* 2.2 Dielectric strength test
2.3 Current output test
2.4 Temperature indication check
2.5 Resistance (conductivity) check
Note: Items marked with an asterisk (*) may only be confirmed by a test certificate.
3.
Inspection Methods, Standards and Conditions
z
Connect the testing circuit as shown in Figure 1. Allow the instrument to warm up for at least
5 minutes before conducting the tests. For the connections for the insulation resistance test,
follow the instructions in Section 3.1 and for the connections for the dielectric strength test,
follow the instructions in Section 3.2.
z
Performance tests should be done in the inspection mode where the tests from Section 3.3
through Section 3.5 take place in sequence and cannot be retraced. If the reconfirmation of
a test is needed, turn off the power to the transmitter, turn on the power again, and enter the
inspection mode to restart the tests.
z
Set the equipment as follows.
Decade resistance box 1 (temperature): 960.9
:
Decade resistance box 2 (conductivity): 10
:
DC source:
24 VDC
3.1 Insulation Resistance Test
(1) Apply 500 V DC between the power supply terminals shorted together (+ and –) and the
earth terminal (G). The insulation resistance must be 100 M
Ω
or greater.
(2) Apply 500 V DC between the input terminals shorted together (11 to 16) and the earth
terminal (G). The insulation resistance must be 100 M
Ω
or greater.
3.2 Dielectric strength test
Apply 600 V AC, an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz
or 60 Hz, between the terminals shown below, for at least 2 seconds. The insulation must
withstand this voltage. (The sensed current should be 10 mA.)
(1) Between the power supply terminals shorted together (+ and –) and the earth terminal (G)
(2) Between the input terminals shorted together (11 to 16) and the earth terminal (G)
(3) Between the input terminals shorted together (11 to 16) and the power supply terminals
shorted together (+ and –)
3.3 Current Output Test
<Test>
(1) Enter Service Code 87 and then password 070.
(2) When the message display shows “*HIF”, press the [YES] key.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...