![YOKOGAWA SC202G User Manual Download Page 42](http://html1.mh-extra.com/html/yokogawa/sc202g/sc202g_user-manual_3799615042.webp)
IM 12D08B02-01E
Installation and wiring 3-9
Extension cable may be purchased in bulk quantities, cut to length. Then it is necessary to terminate the
cable as shown below.
Termination procedure for WF10 cable.
1.
Slide 3 cm of heat shrink tube (9 x 1.5) over the cable end to be terminated.
2.
Strip 9 cm of the outer (black) insulating material, taking care not to cut or damage internal cores.
3 cm
heat shrink
9 cm
remove insulation
Fig. 3-13a.
3.
Remove loose copper screening, and cut off the cotton packing threads as short as possible.
4.
Strip insulation from the last 3 cm of the brown, and the white coaxial cores.
3 cm
cotton threads
Fig. 3-13b.
5.
Extract the coaxial cores from the braid, and trim off the black (low-noise) screening material as
short as possible.
6.
Insulate the overall screen and drain wire (14) and the 2 coaxial screens with suitable plastic tubing.
7.
Strip and terminate all ends with suitable (crimp) terminals and identify with numbers as shown.
11
12
14
15
16
13
17
Red
Blue
Black
White
Brown
Fig. 3-13c.
8.
Finally shrink the overall heat shrink tube into position.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...