IM 12D08B02-01E
Specifications 2-13
2-11. Control Drawing of SC202S FF/PB Specification (FM Intrinsically safe Entity)
Unclassified Location
Classified Location
Division 1
FM Approved
barrier
Voc (Vt)
d
24 V
Ioc (It)
d
250 mA
Poc (Pt)
d
1.2 W
Ca
t
220pF+ Ccable
La
t
0 H + Lcable
Sensor
Connections
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.
FM Class I, DIV. 1, Group ABCD
T4 for ambient temp.
d
55
q
C
-
+
I.S.
certified
Terminator
I.S.
certified
Terminator
SC202S-F
or SC202S-P
Transmitter
Transmitter
-
+
Sensor
Connections
-
+
-
+
x
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or
energy over 20
P
J, or are FM Approvals entity approved and meet connection
requirements.
x
Electrical data of the SC202S-F & SC202S-P:
-
Supply
circuit:
Maximum input voltage Vmax= 24 V
Maximum input current Imax= 250 mA
Maximum input power Pi= 1.2 W
Effective internal capacitance Ci = 220 pF;
Effective internal inductance Li = 0
P
H.
-
Sensor input circuit:
Maximum output voltage Vt= 14.4 V; Maximum output current It= 10 mA
Maximum allowed external capacitance Ca= 59.36 nF
Maximum allowed external inductance La= 340 mH
x
Any FM Approved barrier may be used that meets the following requirements:
Voc or Vt
d
24 V
Ioc or It
d
250 mA
Poc or Pt
d
1.2 W
Ca
t
220 pF + Ccable; La
t
0
P
H + Lcable
When installing this equipment, follow the manufacturer’s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of
Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National
Electrical Code (ANSI/NFPA 70).
Associated apparatus connected to the barrier must not use or generate more than
250 Vrms or Vdc.
x
Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0
Ohm.
x
In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
WARNING
- Substitution of components may impair Intrinsic Safety
- To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing or read, understand and adhere to the manufacturer’s live
maintenance procedures.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...