IM 12D08B02-01E
5-8 Parameter setting
5-3. Service
Codes
Don't set or input service code numbers other than the code numbers defined in this manual. Setting an
undefined service code may make the transmitter malfunction.
When an undefined service code is input by some accident, push the MODE key and escape from the
service level.
5-3-1. Parameter specific functions
Code 01
*SC.RES Choose the required parameter, either conductivity or resistivity. If the
parameter is changed the instrument will go into reset to load parameter
specific default values, followed by starting measurement. For all other service
codes the instrument will return to commissioning mode after the service code
setting is finished.
Code 02
*4.ELEC
Choose the required sensor type. Normally conductivity and/or resistivity
measurements are done with 2-electrode type sensors. At high conductivity
ranges, polarization of the electrodes may cause an error in conductivity
measurement. For this reason 4-electrode type sensors may be necessary.
Code 03
*0.10xC
Enter the factory calibrated cellconstant mentioned on the textplate or on the
fixed cable. This avoids the need for calibration. Any value between 0.008 and
50.0 /cm may be entered. The position of the decimal point may be changed
according the visual description in the right-handed page of section 5-2-2.
*
NOTE
: If the actual cell constant is changed after a calibration or if the entered cell constant differs
from previous value, then the message “*RESET?” will appear on the second line display. After
pressing “YES” the entered value becomes the new nominal and calibrated cell constant. After
pressing “NO” the update procedure of the cell constant entry is canceled.
Code 04
*AIR
To avoid cable influences on the measurement, a “zero” calibration with a dry
sensor may be done. If a connection box (BA10) and extension cable (WF10)
are be used, “zero” calibration should be done including this connection
equipment.
When using a 4-electrode sensor additional connections are required
temporarily Interconnect terminals 13 & 14 with each other and 15 & 16 with
each other before making the adjustment. This is necessary to eliminate the
capacitive influence of the cables. The links should be removed after this step
is completed.
13
14
15
16
Code 05
*POL.CK The EXA SC202 has a polarization check capable of monitoring the signal
from the cell for distortion from polarization errors. If there is a problem with
the installation or the cell becomes fouled, this will trigger E1. For some
application with very low conductivity and long cable runs, this error detection
can cause false alarms during operation. Therefore this code offers the
possibility to disable/enable this check.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...