IM 12D08B02-01E
Specifications 2-1
2-1. Specifications
A. Input specifications
:
Two or four electrodes measurement with
square wave excitation. Cell constants from
0.008 to 50 cm
-1
.
B. Detection method
:
Frequency, read-pulse position and reference
voltage are dynamically optimized.
C. Input ranges
- Conductivity :
Minimum : 0
μ
S/cm
Maximum : 200 mS x (Cell constant)
(overrange 1999 mS / cm).
- Resistivity :
Minimum : 0.005 k
Ω
/ (Cell constant)
Maximum : 999 M
Ω
x cm
- Temperature
Pt1000
: -20 to +250 °C (0 to 500 °F)
Pt100 and Ni100 : -20 to +200 °C (0 to 400 °F)
8K55 NTC : -10 to +120 °C (10 to 250 °F)
PB36 NTC : -20 to +120 °C (0 to 250 °F)
D. Output Span
- Conductivity : - min 0.01
μ
S/cm
: - max. 1999 mS/cm. (max 90%
zero suppression)
- Resistivity : - min 0.001k
Ω
xcm
: - max. 999 M
Ω
x cm. (max 90%
zero suppression)
E. Transmission Signal
: Isolated output of 4-20 mA DC
Burn up (21 mA) or Burn down
(3.6 mA when HART® or distributor comm. is
non-used, 3.9 mA when HART® or distributor
comm. is used) or pulse of 21 mA to signal
failure.
F. Temperature compensation
:
Automatic, for temperature ranges mentioned
under C (inputs).
- Reference temp.
: programmable from 0 to 100 °C or 30 to 210 °F
(default 25 °C).
G. Compensation algorithm
-NaCl
:
According IEC 60746-3 NaCl tables
(default).
-T.C.
:
Two independent user program-
mable temperature coefficients,
from -0.0% to 3.5% per °C (°F)
by adjustment or calibration.
- Matrix
: Conductivity function of concentra-
tion and temperature. Choice out of
mA
mA
mA
mA
5 preprogrammed matrixes and a 25-
point user-programmable matrix.
H. Logbook
:
Software record of important events and diag-
nostic data. Available through HART® link,
with diagnostic information available in the
display.
I. Display
:
Custom liquid crystal display, with a main dis-
play of 31/2 digits 12.5 mm high. Message dis-
play of 6 alpha numeric characters, 7 mm high.
Warning flags and units (mS/cm, k
Ω
·cm,
μ
S/
cm and M
Ω
·cm) as appropriate.
J. Power supply
:
Nominal 24 volt DC loop powered system.
SC202G
; 17 to 40 volts, see Fig.2-1
SC202S
: 17 to 31.5 volts, see Fig.2-2
Maximum load resistance
For the SC202G, see Fig. 1
200
Ω
or less with the PH201G
50
Ω
or less with the SDBT
For the SC202S, see Fig. 2-2
1000
1150
800
600
400
200
150
0
10
20
18
17
30
40
0
Voltage (V)
Load Resistance (
Ω
)
Possible
F06.EPS
Fig.2-1 Supply voltage/ load diagram for the SC202G
0
200
400
600
800
12
16
17
20
24
28
32
425
775
Voltage (V)
Load Resistance (
Ω
)
31.5 V
Possible
Fig.2-2 Supply voltage/ load diagram for the SC202S
K. Input isolation
:
1000
VDC
L. Weight
Body weight
:
approx. 1.6 kg
Mounting brackets weight: approx. 0.7 kg.
2. GENERAL SPECIFICATIONS
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...