IM 12D08B02-01E
Parameter setting 5-9
Code Display Function
Function detail
X Y Z Default values
Parameter specific functions
01
*SC.RES
Select main parameter Conductivity
0
0
Cond.
Resistivity
1
02
*4.ELEC
Select 2/4-EL system 2-Electrode measurement system
0
0
2-El.
4-Electrode measurement system
1
03
0.10xC
Set cell constant
Press NO to step through choice of
1.000 cm
-1
multiplying factors on the second display.
0.10xC
0.10xC
1.00xC
10.0xC
100.xC
0.01xC
Press YES to select a factor
Use >, ^, ENT keys to adjust MAIN digits
RESET?
YES to confirm, NO to cancel
04
*AIR
Zero calibration
Zero calibration with dry cell connected
*START
Press YES to confirm selection
*”WAIT”
Press YES to start, after briefly displaying
*END
“WAIT”, *END will be displayed
Press YES to return to commissioning
mode
05
*POL.CK
Polarization check
Polarization check off
0
1
On
Polarization check on
1
06-09
Not
used
*
NOTE
:
For Code 03: If the actual cell constant is changed after a calibration or if the entered cell
constant differs from previous value, then the message “RESET?” will appear on
the second line display. After pressing “YES” the entered value becomes the new
nominal and calibrated cell constant. After pressing “NO” the update procedure
of the cell constant entry is canceled.
For Code 04: The temperature compensation of NaCl should be selected to confirm zero offset after
*AIR operation.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...