Calibration 6-3
6-3. Calibration with HOLD active
HOLD
ENT
ENT
>
Select the flashing digit with the
>
key.
Increase its value by pressing the key
>
After the indication is stable,
set the value
using the
>
, ,
ENT
key.
Put the sensor in standard
solution. Press
YES
.
When the correct value is displayed,
press
ENT
to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES.
HOLD will be displayed. Press
NO
to turn off
HOLD and return to the measuring mode.