![YOKOGAWA SC202G User Manual Download Page 91](http://html1.mh-extra.com/html/yokogawa/sc202g/sc202g_user-manual_3799615091.webp)
IM 12D08B02-01E
Appendix 3-3
3/3
QIS 12D08B02-01E
Figure 1 Testing Circuit and Test Equipment
SC202G, SC202SJ
SUPPLY SENSOR
+
-
G
11
12
14
13
15
16
24V DC
+
100
:
DC source
+
-
-
DC
Decade
Box 1
Note 1: Cable connected to sensor input should be conductivity detector cable of
length 2.1 ±0.1 m. Connect pins 13 and 14, also 15 and 16, to Decade box 2
terminals. Use shielded cable, and connect shield to pin 14.
Note 1
Decade
Box 2
Ground
Milli-
ammeter
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...