IM 12D08B02-01E
Specifications 2-15
2-12. Control Drawing of SC202S FF/PB Specification (FM Intrinsically safe FISCO)
-
+
SC202S-F
or SC202S-P
Transmitter
-
+
Transmitter
Unclassified Location
Classified Location
Division 1
FM Approved
FISCO barrier
Voc (Vt)
d
17,5 V
Ioc (It)
d
380 mA
Poc (Pt)
d
5,32 W
Sensor
Connections
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.
FM Class I, DIV. 1, Group ABCD
T4 for ambient temp.
d
55
q
C
Sensor
Connections
-
+
-
+
FM Approved
Terminator
R = 90..100
Ω
C = 0..2,2
μ
F
FM Approved
Terminator
R = 90..100
Ω
C = 0..2,2
μ
F
x
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which
neither store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or
energy over 20
P
J, or are FM Approvals entity approved and meet connection
requirements.
x
Electrical data of the SC202S-F & SC202S-P:
-
Supply circuit: Vmax = 17.5 V; Imax = 380 mA; Pi = 5.32 W; Ci = 220 pF; Li = 0
P
H.
-
Sensor input circuit: Vt = 14.4 V; It = 10 mA; Ca = 59.36 nF; La = 340 mH
x
Any FM Approved FISCO barrier may be used that meets the following requirements:
Voc or Vt
d
17.5 V; Ioc or It
d
380 mA; Poc or Pt
d
5.32 W
When installing this equipment, follow the manufacturer’s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of
Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National
Electrical Code (ANSI/NFPA 70).
Associated apparatus connected to the FISCO barrier must not use or generate more
than
250
Vrms or Vdc.
x
Resistance between FISCO Intrinsically Safe Ground and earth ground must be less than
1.0 Ohm.
x
In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
x
The FISCO concept allows the interconnection of several I.S. apparatus not specifically
examined in such combination. The criterion for such interconnection is that the voltage
(Vmax), the current (Imax) and the power (Pi) which I.S. apparatus can receive and remain
intrinsically safe, considering faults, must be equal to or greater that the voltage (Voc, Vt),
the current (Ioc, It) and the power (Poc, Pt) which can be providede by the FM approved
FISCO barrier. In addition, the maximum unprotected residual capacitance (Ci) and
inductance (Li) of each apparatus (other than the terminator) connected to the Fieldbus
must be less than or equal to 5nF and 10
μ
H respectively.
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...