IM 12D08B02-01E
5-18 Parameter setting
5-7. Communication setup
Code 60
*COMM.
The settings should be adjusted to suit the communicating device connected to
the output. The communication can be set to HART® or to PH201*B distributor
(see Appendix 2).
*ADDR.
Select address 00 for point to point communication with 4-20mA transmission.
Addresses 01 to 15 are used in multi-drop configuration (fixed 4mA output).
Code 61
*HOUR
The clock/calendar for the logbook is set for current date and time as reference.
*MINUT
*SECND
*YEAR
*MONTH
*DAY
Code 62
*ERASE
Erase logbook function to clear the recorded data for a fresh start. This may
be desirable when re-commissioning an instrument that has been out of service
for a while.
5-8. General
Code 70
*LOAD
The load defaults code allows the instrument to be returned to the default set
up with a single operation. This can be useful when wanting to change from
one application to another.
5-9. Test and setup mode
Code 80
*TEST
Not used.
NOTE
: Attempting to change data in service code, 80 and above without the proper instructions and
equipment, can result in corruption of the instrument setup, and will impair the performance of
the unit.
mA
mA
Summary of Contents for SC202G
Page 8: ...IM 12D08B02 01E LT LV PL EST SLO H BG RO M CZ SK...
Page 13: ...IM 12D08B02 01E 1 4 Introduction...
Page 33: ...IM 12D08B02 01E 2 20 Specifications...
Page 74: ...IM 12D08B02 01E Spare Parts 10 1 10 SPARE PARTS See Customer Maintenance Parts List...
Page 92: ...IM 12D08B02 01E 3 4 Appendix...
Page 96: ...IM 12D08B02 01E 3 8 Appendix...
Page 99: ...IM 12D08B02 01E Appendix 3 11 3 3 QIS 12D08B02 61E Figure 1 Testing Circuit and Test Equipment...
Page 100: ...IM 12D08B02 01E 3 12 Appendix...
Page 104: ...IM 12D08B02 01E 3 16 Appendix...