<6. SETTING AND CALIBRATION>
6-35
IM 11G02N01-01E
Description of Calibration Log screen
Caution
• Calibration Log screen
ENT
F6_8_1.ai
Past calibration history.
Sensor input value, concentration value, and
the date when zero/span calibration is performed
are logged. The 10 newest calibration data is
logged by each component.
Move the cursor to Clear Calibration Log and
press the
ENT
key, and the calibration log is
cleared completely.
Z1 : Zero calibration (Z) of Range 1
S1 : Span calibration (S) of Range 1
M : Value of measuring detector at the time
of calibration
C : Value of the interference compensation
detector at the time of calibration
Con : Concentration value displayed before
calibration
If the following operation is maladjusted, the measurement may be adversely and excessively
affected. Carry out the operation with utmost attention.
• Optical adjustment screen
F6_8gamen.ai
For details of this item, refer to “Item 7.3.3 Optical
zero adjustment method”.
Press
ENT
key and turn ON the solenoid valve
signal for each calibration gas by using the
or
key.
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...