<6. SETTING AND CALIBRATION>
6-9
IM 11G02N01-01E
6.2.4 Setting of auto calibration component/range
Select the Ch (component) and the range with which auto calibration is to be performed. The Ch for
which “AR” has been selected as range switch mode is calibrated in the range set here even when
auto calibration is performed.
MODE
ENT
( )
ENT
( )
ENT
( )
F6_2_4.ai
(1) During measurement, press the
MODE
key
to display the User Mode.
(2) Point the cursor to “Calibration
Parameters” by pressing the
or
key. Press the
ENT
key.
(3) In the “Calibration Parameters” screen
that appears, point the cursor to “Auto
Calibration Components / Range” by
pressing the
or
key. Press the
ENT
key.
(4) In the “Auto Calibration Components /
Range” selection screen that appears, point the
cursor to the Ch you want to set by pressing
the
or
key. Press the
ENT
key.