<8. TROUBLE SHOOTING>
8-2
IM 11G02N01-01E
In case of Error No. 5 and No. 7
1 3 6
• Pressing delete the error display.
Calibration is continued. Unless another calibration
error occurs, calibration is carried out to the end, the
Measurement screen returns.
C H
O
2
0-25
vol%
0 0 9
.
C H
CO
0-100
ppm
0 0
C H
CO
2
0-10
vol%
0 0 0 0
C H
SO
2
0-100
ppm
C H
NO
2
0-25
ppm
9 0 8
.
.
.
ENT
ENT
ESC
.
ESC
Calibrated
forcedly
F8_2.ai
Error log file
If error occurs, the history is saved in an error log file. The error log file exists in the maintenance
mode.
Error log screen
Errors that occurred
New
Date and time
when an error occurred.
Component with which
the error occurred.
Old
F8_3.ai
* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every time a new occurs.
* If the power supply is turned OFF, the contents in the error log file will not be lost or damaged.
Deletion of error history
Press the
ENT
key on the above screen, and the “Error Log Clear” will be inverted. Further pressing
the
ENT
key will clear the error history.