<3. INSTALLATION>
3-10
IM 11G02N01-01E
3.5 Wiring
3.5.1 Power inlet
The power inlet is provided at the rear panel.
Connect supplied power cable to this power inlet.
Input/Output terminal connector
Power inlet
Rear panel
Grounding 2-pole plug
F3_5_1.ai
● Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder)
that generates much electrical noise. If using the
instrument near such a noise generating unit is
unavoidable, use a different power line to avoid noise.
● Mount a noise suppressor such as varistor or spark
killer as shown at right figure to the noise generating
unit when noise is generated from relays or solenoid
valves. Mount the suppressor near the noise generating
source, or it will have no effect.
3.5.2 Input/Output terminal module
This analyzer should be connected to the input/output terminal module by supplied exclusive cable.
Plug this cable connector into the receptacle at the rear panel of the analyzer and the receptacle on
the PC board of the input/output module.
Connect the exclusive cable so that the ferrite core attached to the cable comes to the analyzer side.
Exclusive cable
(1 meter long)
Analyzer
Input/Output
terminal module
Ferrite core
(to analyzer side)
F3_5_2.ai
Main unit
power supply
Varistor or
spark killer
Install (connect)
near the source.
Noise
generating
source
F3_512.ai
When noise source is in the vicinity
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...