<6. SETTING AND CALIBRATION>
6-30
IM 11G02N01-01E
(6) You can select the value for hold from the value immediately before entering output hold, “current,”
and arbitrary value, “setting.”
Follow the procedures shown below to make the setting.
ENT
ENT
( )
F6_7_5.ai
1) Press the
ENT
key in a state where the
cursor is placed next to Hold.
2) “ON” or “OFF” is highlighted. Press the
or the
key to select ON or OFF.
Press the
ENT
key to return to (1).
3) Press the
key in a state ON/OFF is
highlighted, and “Current” or “Setting” is
highlighted. Select “Current” or
“Setting” by pressing the
or the
key.
4) Press the
ENT
key while “Current” is
selected to return to (1). Press the
ENT
key while “Setting” is selected to go to
the setting entering screen.
“Current”: Holds the value immediately
before the hold.
“Setting”: Holds the value arbitrarily set.