<6. SETTING AND CALIBRATION>
6-36
IM 11G02N01-01E
Description of moisture interference
adjustment screen
Caution
• Moisture interference adjustment screen
For details of this item, refer to “Item 7.3.4
ENT
F6_8_2.ai
Moisture interference adjustment method.”
In values on the left side of screen, the moisture
interference for each component is already offset.
The figures at right are interference compensation
coefficients.
Move the cursor to a desired Ch (component)
by pressing the
or the
key, and then
press the
ENT
key, and the selected value at right
is highlighted.
Check that the gas for moisture interference
compensation is flowing, change the moisture
interference compensation coefficient using
the
or the
key, adjust the value at left
so that it becomes near zero, and then press
the
ENT
key to log moisture interference
compensation value.
Since an interference compensation detector
is not provided if the 1st range is beyond 0 to
10 vol%, no interference adjustment can be
performed (no need).