<6. SETTING AND CALIBRATION>
6-12
IM 11G02N01-01E
Cursor for setting value
ENT
End of Alarm Setting
F6_3_1_2.ai
(5) After setting, the alarm setting is now
completed by pressing the
ENT
key.
To close the “Alarm Setting” or to cancel this
mode midway, press the
ESC
key.
A previous screen will return.
0% to 100%FS (Settable in each range).
The alarm contact assigned the same number as the alarm is operated accordingly.
Channel:
Channel setting targeted for issuance of alarm (Power off alarm can be selected
for alarm 6.) One Ch No. can be selected for multiple alarms.
H-Limit value: Sets the high limit value (concentration) of alarm.
L-Limit value: Sets the low limit value (concentration) of alarm.
Kind of Alarm: Selects one of High limit alarm, Low limit alarm, and High limit or Low limit alarm,
HH limit alarm, and LL limit alarm.
High, HH ... Alarm contact closes when above H-limit alarm.
Low, LL ... Alarm contact closes when below L-limit alarm.
High or Low ... Alarm contact closes when above H-limit value or below lower limit
value.
If “Power” is selected for Channel, the contact is closed at all times while the power is on irrespective
of the setting made here. (Alarm-6 only)
ON/OFF: Enables the alarm function if set at ON, or disables the alarm function if set at OFF.
●The H-limit value cannot be set below the L-limit value, and the L-limit value cannot be set above the H-limit value.
If it is desired to set the H-limit value below the L-limit value already stored in the memory, reduce the L-limit value
beforehand, and vice versa.
C h
O
2
0-25
vol%
C h
CO
0-100
ppm
C h
CO
2
0-10
vol%
C h
SO
2
0-100
ppm
C h
NO
X
0-100
ppm
0 0
.
0 0 0 3
.
0 0
.
2 1 0 0
.
- - - -
H-alarm
F6_3_1_3.ai
When an H-limit alarm occurs, the “H-alarm”
message comes on in the field of relevant
Ch (component).
(“L-alarm” for L-limit alarm, “HH-alarm”
for HH-limit alarm, and “LL-alarm” for
LL-limit alarm)
For 10 minutes after turning on power, the alarm judgment is inactive.
To close the “Alarm Setting”
Setting range
Description of setting items
Typical on-screen display when an alarm occurs
Note
Summary of Contents for IR400 series
Page 1: ...User s Manual R Model IR400 Infrared Gas Analyzer IM 11G02N01 01E IM 11G02N01 01E 5th Edition ...
Page 7: ...Blank Page ...
Page 11: ...Blank Page ...
Page 33: ...Blank Page ...
Page 35: ...Blank Page ...
Page 83: ...Blank Page ...
Page 93: ...Blank Page ...
Page 97: ...Blank Page ...
Page 117: ...Blank Page ...