1-3
IM DM7560-01EN
Name and function of each part
1
1.1.2 Input terminal part
Figure 1.3 shows the input terminal part on the front panel and Table 1.3 describes the name and
function of each part.
1
2
3
5
4
Figure 1.3 Input terminal part
Table 1.3 Input terminal part
№
Name
Contents and functions (outline)
1
INPUT V · Ω ·
HI-LO input terminal
It is the input terminal for measurement of voltage (DCV, ACV), resistance (2WΩ),
temperature (TEMP), diode ( ) and Continuity test (CONT). The attached test lead (pair
of red and black leads) or banana terminal is connected to it. Take care of the range and
maximum permissible input.
2
SENSE 4WΩ
HI-LO input terminal
It is used for resistance measurement (4WΩ) and temperature measurement (TEMP,
RTD-4Wire).
• Max. permissible input: 200 Vpeak for all ranges
3
I-LO
input terminal
It is the input terminal for current measurement (DCI, ACI).
• Max. permissible input : 3 A DC or rms (continuous)
/ 250 V (Open circuit voltage)
4
FUSE
The fuse is installed for overcurrent protection when measuring current (DCI, ACI).
• Fuse specification: F3A, 250 V
In addition to the fuse installed in this instrument, two fuses are attached.
5
POWER switch
It is the power switch of this instrument.
• ON: I (switch is pressed)
• OFF:
(switch is not pressed)
1.1 Name and outline of each part on front panel