4-9
IM DM7560-02EN
Specifications
4.2.5 2 terminal resistance measurement (2WΩ)/4 terminal resistance measurement
(4WΩ)
4.2.5.1 Resolution, accuracy and measuring current
Unit: ±(% of % of range)
Range
Full scale
Resolution
Accuracy
Temp. coefficient
Measuring current
100 Ω
119.9999
0.1 mΩ
0.010 + 0.004
0.0006 + 0.0005
Approx. 1 mA
1k Ω
1.199999
1 mΩ
0.010 + 0.001
0.0006 + 0.0001
Approx. 1 mA
10k Ω
11.99999
10 mΩ
0.010 + 0.001
0.0006 + 0.0001
Approx. 100 μA
100 kΩ
119.9999
0.1 Ω
0.010 + 0.001
0.0006 + 0.0001
Approx. 10 μA
1 MΩ
1.199999
1 Ω
0.010 + 0.001
0.0010 + 0.0002
Approx. 5 μA
10 MΩ
11.99999
10 Ω
0.040 + 0.001
0.0030 + 0.0004
Approx. 500 nA
100 MΩ
119.9999
100 Ω
0.800 + 0.010
0.1500 + 0.0002
Approx.500 nA //10 MΩ
• Sampling rate: 1 S/s
• This is accuracy by 6 and 1/2 digits resolution for 4 terminal resistance measurement or 2
terminal resistance measurement after zero compensation by NULL calculation. In the case
NULL calculation is not performed, 0.2Ω additional error is added to 2 terminal resistance
measurement.
• Maximum allowable voltage
Between Ω-COM terminals: 800 Vpeak (continuous), or 1100 Vpeak (for 1 minute)
Between Sense Hi-Lo: 200 Vpeak
• Terminal open voltage <17 V
4.2.6
Continuity test (
)
4.2.6.1 Accuracy, resolution and measuring current
Unit: ± (% of r % of range)
Resist. range
Resolution
Threshold
Accuracy
Temp.
coefficient
Measuring
current
Sampling rate
1 kΩ
10 mΩ
1 Ω to 1000 Ω
0.010 + 0.020 0.001 + 0.002
Approx.1 mA
100 S/s
• Sounding of electronics buzzer
• Maximum allowable voltage : 800 Vpeak (continuous), or 1100 Vpeak (for 1 minute)
4.2.7 Diode ( )
4.2.7.1 Accuracy and measuring range
Unit: ± (% of r % of range)
Measuring
current
Measuring range
Accuracy
Temp. coefficient
Terminal open
voltage
Sampling rate
Approx.1 mA
0.01 mV to 1.19999 V 0.010 + 0.020
0.001 + 0.020
<17 V
100 S/s
• Maximum allowable voltage : 800 Vpeak (continuous), or 1100 Vpeak (for 1 minute)
4.2 Specifications of Basic Measuring Function