4-12
IM DM7560-02EN
4.3
Trigger function
Trigger mode
AUTO
Measures automatically in accordance with sampling rate and interval.
SINGLE
Measures in accordance with TRIG input.
Trigger source
Back TRIG input terminal
Polarity and valid/invalid are switchable by menu.
(External trigger)
HOLD/TRIG key
Enters by key manually.
(Manual trigger)
REMOTE
Please refer to the Comunication Interface User's Manual (PDF files
are stored in CD)
Trigger sample number
Sets the number of continuous data measurement per one trigger.
Setting range
1 to 100000
Trigger delay
Sets delay time from TRIG input to data measurement.
Setting range
0.00 ms to 3600 s
Resolution
10 μs
Interval
Measuring interval setting of sampling
*This is valid when larger value than current sampling rate is set.
Setting range
0.00 ms to 3600 s
Resolution
10 μs