4-10
IM DM7560-02EN
4.2.8
Temperature measurement (TEMP, TC: THERMOCOUPLE)
4.2.8.1 Accuracy and resolution
Unit: ± (% of r °C)
Thermocouple
Measuring range (°C)
Accuracy
Resolution
Max. allowable voltage
R
−50 to 0
0.20 + 0.70
0.001°C
800 Vpeak
(continuous),
1100 Vpeak
(for 1 minute)
0 to +100
0.20 + 0.50
+100 to +1765
0.20 + 0.30
K(CA)
−200 to −100
0.15 + 0.50
−100 to 0
0.15 + 0.35
0 to +1370
0.15 + 0.20
T(CC)
−200 to −100
0.15 + 0.50
−100 to 0
0.15 + 0.35
0 to +400
0.15 + 0.20
J(IC)
−200 to −100
0.15 + 0.50
−100 to 0
0.15 + 0.35
0 to +1200
0.15 + 0.20
E(CRC)
−200 to −100
0.15 + 0.50
−100 to 0
0.15 + 0.35
0 to +1000
0.15 + 0.20
• Sampling rate: 1 S/s
• Thermocouple accuracy is not included.
• Cold junction temperature shall be input by TEMP/SENSOR menu and does not include its
error.
• In calculational guarantee temperature 0°C to 18°C and 28°C to 50°C, ±0.1°C/°C is added to
all thermocouples.
• It is outside the accuracy guarantee though temperatures that are lower than -200°C might be
displayed as measurements.
• Standard heat electromotive force depends on line graph approximate calculation by JIS C
1602.
• This instrument does not support internal reference junction compensation. Set the reference
junction compensation temperature manually (see section 4.3.9 in the IM DM7560-01EN).
Internal reference junction compensation: A function that measures internally the input
terminal temperature and compensates the temperature at the measurement point.
4.2.9
Temperature measurement (TEMP, RTD: Resistance
temperature detector)
4.2.9.1 Measuring range, accuracy and resolution
RTD
Measuring range (°C)
Accuracy
Temp. coefficient
Resolution
Pt100
−200 to +850
±
0.06°C
±
0.003°C/°C
0.001°C
JPt100
−200 to +510
• Sampling rate: 1 S/s
• This complies with Pt100 : JIS C1604 standards.
• This complies with JPt100 : JIS C1604 standards.
• In 4 lead wire system, accuracy of measuring cable (or probe) is not included.
• Maximum allowable voltage : 800 Vpeak (continuous), or 1100 Vpeak (for 1 minute)
4.2 Specifications of Basic Measuring Function