IM CA71-E
8-9
Communication Function
1
2
3
4
5
6
7
8
9
10
11
12
13
App
SY
Syntax for setting
SYm<delimiter>
Syntax for query
SY?<delimiter>
⇒
Response: SYm<delimiter>
Description of parameter
m: Mode
m=0: Normal mode
m=1: Adjustment mode
CD
When adjustment
When adjustment
When adjustment
When adjustment
Syntax for setting
CDm<delimiter>
Syntax for query
CD?<delimiter>
⇒
Response: CDm<delimiter>
Description of parameter
m: Sourced setpoint (8 digits) ex. +1.00003
CL
Syntax for setting
CLm<delimiter>
Syntax for query
CL?<delimiter>
⇒
Response: CLm<delimiter>
Description of parameter
m: Adjustment item
m=3: Adjustment of source
m=4: Adjustment of measurement
CP
Sets the adjustment point.
Syntax for setting
CPm<delimiter>
Description of parameter
m: Adjustment point
m=0: FS adjustment
m=1: Zero adjustment
CW
Saves the adjusted data.
Syntax for setting
CW<delimiter>
When normal
condition/adjustment
Switches between the normal and adjustment
modes/queries the current setting.
Sets the sourced setpoint/queries the current setting.
Sets the adjustment item/queries the current setting.
Be sure to execute CW command after adjustment for each function/range.
Without executing CW command, the adjusted value will be deleted when
the power is turned off.
8.7 Detailed Description of Commands