4-14
IM CA71-E
4.5 Sourcing Pulse Signals
You can source a preset type of continuous pulse train,
a pulse signal with a preset frequency, or the preset number of
pulses.
Frequency-based
signal
Source of
number of pulses
n = Preset number
of pulses
Continued
Amplitude setpoint
0V
OFF ON
1
2
3
n
Automatically turned off
Providing Pulse Output
Press key
4.5.1 Sourcing a Continuous Pulse Train
Step 1:
Using the function selector switch, select
PULSE
.
The LCD shows the default frequency
.
Step 2:
Using the
key, set the frequency range.
Pressing of the
key cycles through the 500.0 Hz,
1000 Hz and 10 kHz options.
Step 3:
Set the output value digit by digit using each pair of
and output setting keys.
Each pair of and keys corresponds to each digit of
the LCD reading.
Each press of the or key increases or decreases
the digit.
Increasing the digit from 9 or decreasing it from 0 causes
the digit to overflow or underflow, allowing you to set
the output value without interruption.
Holding down the or key continuously changes
the digit in question.
Pressing the
key initializes the output setpoint to
the default value (differs depending on the frequency range).
Step 4:
Pressing the
key once switches to amplitude setting
mode.
The LCD provides a reading of .
4.5 Sourcing Pulse Signals