3. Before Starting Source/Measurement
NOTE
• Use the instrument under the following environmental conditions if precise
source or measurement is your requirement:
Ambient temperature range: 23±5°C; ambient humidity range: 20 to 80% RH
(non-condensing)
When using the instrument within a temperature range of 0 to 18°C or
28 to 50°C, add a value based on the temperature coefficient shown in
Chapter 12, “Specifications (page 12-1),” to the given accuracy rating.
• When using the instrument at an ambient humidity of 30% or lower, prevent
electrostatic charges from being produced, by using an antistatic mat or any
other alternative means.
• Condensation may occur if you relocate the instrument from places with
low temperature and humidity to places with high temperature and humidity,
or if the instrument experiences any sudden temperature change.
In that case, leave the instrument under the given ambient temperature for
at least one hour to ensure that the instrument is free from condensation,
before using the instrument.
■ Installing or Replacing the Batteries
WARNING
To avoid electrical shock, always remove the source or measurement lead
cables from the object under test, as well as from the instrument itself.
French
AVERTISSEMENT
Pour éviter tout risque d'électrocution, retirer toujours les câbles de la source
ou du cordon de mesure de l'objet testé, ainsi que de l'instrument lui-même.