IM CA71-E
7-5
Functions Provided by DIP
Switch
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Step 8:
Using each pair of and keys, set the value of
the denominator
m
and the starting setpoint of
the enumerator
n
.
(See Section 4.6, "Divided Output Function (
n/m
),
for further details.)
The starting setpoint is the minimum of the variable
n
for
auto step operation.
Step 9:
Pressing the
key initiates the automatic stepping of
the divided output (
n/m
), as described below.
Assuming the starting value of the variable
n
is 1,
the output cyclically changes with the variable
n
as
n
changes
in the following manner.
n
= 1 → 2 → 3 → · · · (
m
– 1) →
m
→ (
m
– 1) → · · · →
2 → 1 → 2 → 3 → · · ·
Step 10:
To quit the auto step function, turn off the calibrator by
pressing the
key.
Place switch 1 back in the OFF (left-side) position.
Note that disabling the divided output mode (
n/m
) before
placing switch 1 back in the OFF position enters sweep mode.
TIP
• Pressing the
key to change to the SOURCE OFF state pauses
the auto step operation.
• To execute the auto step function again, press the
key.
The auto step operation resumes from the value of n shown on the LCD.
7.2 Auto Step Function