IM CA71-E
7-7
Functions Provided by DIP
Switch
1
2
3
4
5
6
7
8
9
10
11
12
13
App
7.4 Selecting the IPTS-68 Function
By placing switch 4 (IPTS-68 switch) in the ON (right-side) position,
you can select the IPTS-68 temperature scale when you choose
the type-K, E, J, T, N, R, S or B thermocouple or the Pt100 RTD.
Placing the switch in the OFF position results in the selection of
the ITS-90 temperature scale.
TIP
When you have selected a type of thermocouple or RTD,
the LCD shows IPTS-68.
7.5 Switch Not Used
Although switch 5 (No Use switch) of the DIP has no effect on
calibrator operation, the switch should be placed in the OFF
(left-side) position.
7.6 Temp Switch
Check that the switch should be placed in the OFF position.
7.7 Selecting the Contact In Function
(Contact Input for Pulse Measurement)
By placing switch 7 (Contact In switch) in the ON (right-side) position,
you can measure transistor contact on-off signals.
If you select the contact input function, the (contact) symbol
appears on the LCD.
By placing the switch in the OFF position, you can measure normal
pulse.
7.4 Selecting the IPTS-68 Function