5-11
IM AQ6380-02EN
Maintenance
1
2
3
4
5
6
App
Index
No. Message
Cause of Warning
180 APP: File not found
There is no valid update file for installing an application.
181 APP: File format error
The application update file format is incorrect.
182 APP: File data error
The application update file data is incorrect.
183 APP: Out of storage space
There is not enough free space to install the application.
184 APP: File operation error
A file operation error occurred during application installation.
185 APP: Unexpected error
An unexpected error occurred during application installation.
No. 200 - 299: Warnings for hardware failures
200 Fan motor stopped
The fan motor (main unit) stopped. Automatically shuts down after
10 seconds from occurrence.
201 Calibration data failed
Started up in emulation mode because there was a problem with
the instrument’s calibration data.
204 Calibration data failed
There is an error in the instrument’s calibration data.
205 Internal communication error
An abnormality occurred during instrument-internal communication.
206 Internal communication error
An abnormality occurred during instrument-internal communication.
207 Abnormal internal temperature
Internal temperature is abnormally high. Automatically shuts down
after 10 seconds from occurrence.
210 Internal Temparature warning
Warning for rising internal temperature.
211 Auto offset error
Obtained an abnormal value during AUTO OFFSET operation.
212 Auto offset error
Obtained an abnormal value during AUTO OFFSET operation.
213 Auto temperature control error
Problem with the light detector temperature control.
214 Measurement sequence error
The sweep stopped because measurement sequence fell into
disorder during the sweep.
215 System optimization required
System optimization is required. Restart the system.
220 Boot sequence error
Started up in emulation mode because an abnormality occurred
upon start-up
221 Boot sequence error
Started up in emulation mode because an abnormality occurred
upon start-up
222 Emulation mode.
Entered emulation mode because an abnormality occurred.
223 Boot sequence error!
Started up in emulation mode because an abnormality occurred
upon start-up
224 Internal communication error
An abnormality occurred during instrument-internal communication.
225 Internal communication error
An abnormality occurred during instrument-internal communication.
226 Internal communication error
An abnormality occurred during instrument-internal communication.
227 Internal communication error
An abnormality occurred during instrument-internal communication.
228 Memory allocation error
Data memory initialization failed.
230 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
231 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
232 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
233 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
234 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
235 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
236 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
237 Monochromator error
Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
5.7 Warning Display Function