3.1 Installing the Instrument
Ambient Temperature and Humidity
Ambient temperature: 5 to 35°C
Ambient humidity:
20%RH to 80%RH (no condensation present)
Note
Condensation may occur if the instrument is moved to another place where the ambient temperature is
higher, or if the temperature changes rapidly. In such cases, allow sufficient time for the instrument to adjust
to the ambient temperature before use.
When the instrument is packaged in a box and moved, prevent condensation by allowing sufficient time for
the instrument to acclimatize before removing it from the box.
Flat, Even Surface
Install the instrument with the correct orientation on a stable, horizontal surface. If the instrument
is installed in a horizontal position, rubber stoppers can be attached to the feet at the rear of the
instrument to prevent the instrument from sliding. One set of rubber stoppers (two stoppers) are
included with the instrument.
Rubber stoppers
A9088ZM
Feet at the rear of the instrument