Using the OSICS DFB Module
94
OSICS User Manual
8.
Press the knob to confirm.
The message
Moving...
appears while the power value is set. The power calibration
is now complete and the values are stored in the module's non-volatile memory.
The DFB
Unit Setup
menu is displayed.
The new calibration data sets the new system calibration default.
8.7.3
Asking for a Factory Recalibration
Subject
To externally verify the performance of the OSICS, refer to the procedure described in
Performing a Wavelength Calibration, p. 92
and section
. If the performance does not meet warranted specifications, a complete
calibration of the instrument may be required.
We recommend performing factory calibration of OSICS DFB modules over one-year
intervals.
Procedure
Contact the
Y
enista Optics
customer support service (see section
) for a complete calibration, as some calibration steps require the adjustment of
internal components including the laser head.