Using the OSICS T100 Module
OSICS User Manual
199
Figure 152: T100 Unit Setup –
λ
/f calibration, first wavelength
This is the first step of the two-point wavelength/frequency calibration method.
The first wavelength/frequency is the first factory calibration wavelength/frequency.
The value you enter in this field is the wavemeter reading.
5.
Enter the value measured at the wavemeter as follows:
a. Turn the rotary knob to put the cursor under the digit you want to modify and
press the knob to highlight it.
b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and
press the knob to validate the selected digit.
c. Perform steps a. and b. again with the other digits to modify.
d. Turn the rotary knob clockwise to put the cursor under
ENTER
and press the
knob to validate the value.
The message
Moving...
appears while the first wavelength/frequency is being
set.
The
Wavelength/Frequency calibration
window appears, showing the second
wavelength value.
Figure 153: T100 Unit Setup –
λ
/f calibration, second wavelength
This is the second step of the two-point wavelength/frequency calibration method.
6.
Measure the second wavelength/frequency using the wavemeter.
7.
Enter the wavelength/frequency reading in this field and select
ENTER
.
A confirmation message appears.
8.
Press the knob to confirm.
While updating the emission wavelength/frequency, the system displays the
Moving…
message until the wavelength/frequency is set.
Main
EXIT
2
3
4
5
6
7
8
ECL Unit Setup - Wavelength calibration
ESCAPE
First wavelength value :
Correct the value.
ENTER
1
▲
λ
= 1540.461 nm
Main
EXIT
2
3
4
5
6
7
8
ECL Unit Setup - Wavelength calibration
ESCAPE
Second wavelength value :
Correct the value.
ENTER
1
▲
λ
= 1589.694 nm