Using the OSICS ECL Module
110
OSICS User Manual
9.7
Configuring the OSICS ECL Signal Modulation
Presentation
In addition to the OSICS Mainframe built-in modulation source (see section
the OSICS Mainframe Modulation, p. 40
), the OSICS ECL module holds a built-in digital
modulation source.
The OSICS ECL module supports the following modulation schemes:
•
Internal modulation:
•
Digital (TTL) modulation using the OSICS ECL internal source
•
Digital (TTL) modulation using the OSICS Mainframe internal source
•
External modulation:
•
Digital (TTL) modulation using an external source connected to the OSICS
Mainframe
•
Analog modulation using an external source connected to the OSICS ECL
The
OUT 2
BNC connector can be used as an external modulation source for other
devices.
Modulation
Scheme
Figure 71: Mainframe and ECL Module Modulation Scheme
26&,//$725
26,&60DLQIUDPH
0RGLQ
%1&&RQQHFWRU
6\QFKUR
%1&&RQQHFWRU
0RGXODWLRQ
,QSXW
0RGXODWLRQ
2XWSXW
(;7(51$/
,17(51$/
0RGLQ
/('
0$,1)5$0(
+]²0+]
26&,//$725
+]²0+]
,17(51$/
26,&67PRGXOH
26,&6(&/PRGXOH
287
%1&&RQQHFWRU
0RGXODWLRQ
2XWSXW
2))
21
21
,19(57('
60%&RQQHFWRU
$QDORJ
0RGXODWLRQ
,QSXW
77/
0RGXODWLRQ
/DVHU
'LRGH